|
|
|
||||||||||||||||||||
|
|
Power Supply Group definitions in medically used Rooms according IEC 60364-7-710
|
710.3.5 Group 0 |
710.3.6 Group 1 |
710.3.7 Group 2 |
Medical location where no applied parts are intended to be used |
Medical location where applied parts are intended to be used as follows:
|
Medical location where applied parts are intended to be used in applications such as intra-cardiac procedures, operating theatres and vital treatment where discontinuity (failure) of the supply can cause danger of life |
|
|
|
![]() |
![]() |
![]() |
Medical IT System in Group 2 Locations
In group 2 medical locations, the medical IT system shall be used for circuits supplying
- Medical electrical equipment
- Systems intended for life support
- Surgical applications
- Other electrical equipment located in the "Patient Environment"
The IT system is supplied by an independent power source, e.g.
- Distribution / isolation transformer
- Control-power transformer
- Generating set
- Battery
- Inverter of an UPS
The IT system has no direct connection between live conductors and earth
- All exposed conductive parts of the electrical installation are earthed
- IEC 60364-7-71:A permanent insulation monitoring device must be designed to indicate the first occurrence of a live mass or earth fault; it must trigger an audible or visual signal.
Need for Medical IT System
Insulation Fault
- In the event of an insulation fault RF only a very low current ICe flows
- Fuse does not trip
- In the event of a single-pole earth fault, the power supply is guaranteed
- No interruption of operation
- An Alarm is indicated if an A-ISOMETER is installed
Leakage Current
- The IT system is a small local network with low leakage capacitances.
- The fault current is limited by the body impedance, the earthing resistance and the high impedance of the fault loop.
- Hazards to persons and equipment arising from high fault currents are reduced.
Fire Risk
- A small current flows limited by the high impedance of the fault loop.
- Considerable less fire risk
- Increased protection for persons and equipment
REQUIREMENTS OF MEDICAL IT SYSTEM Protection against Dangerous Overloads and Insulation Failure Requirements for Power Supply Transformer
|
![]() ![]() ![]() ![]() |
Monitoring of Transformer
- Monitoring of overload and high temperature for the medical IT transformer is required
- But no circuit breaker for protection against overload and overcurrent should be used (risk: unexpected tripping of the fuse)
- Overcurrent protective devices (MCB) should only be used for short-circuit protection.
- If too many medical electrical devices are in use in an operating room, the information will be displayed on a remote indicator if provided.
Requirements for Insulation Monitoring Device
Hence the essential requirements for Medical IT System are Monitoring of Load, Temperature and Insulation |
![]() |
![]() |
Requirements for Remote Alarm Indications
|
Insulation Fault Location
- The problem
- In intensive care locations several socket-outlets are arranged at every patient place, sometimes up to 24 socket outlets and more
- In an intensive care unit with 4 beds, for example, these are more than 90 socket outlets
- In intensive care units usually no technical experts are present, but medical personnel
- If a defective device causes an insulation fault, it will be very difficult to locate the associated socket outlet
- In order to guarantee the patient's safety, the insulation fault must be found as quick as possible.
- The solution
- Insulation fault location system EDS474
Functional Diagram of EDS System
Benefits
- Ungrounded Power Systems according to IEC364-7-710
- Fault location systems offer precise and fast localisation of defective electrical equipment offer and guarantee a higher protection for patients and staff
- Modern and sophisticated indication panels and visualization possibilities offer precise and fast information to medical and technical staff
- Testing the electrical safety of medical electrical equipment acc. to IEC 62353 / 60601-1









